Functional test log
SonoCheck ultrasonic cavitation monitor
Date:
Ultrasonic cleaner:
Program / time:
.
.
.
Prepare the ultrasonic cleaning bath as per
manufacturers instructions. Make sure
water level and degassing procedures have been followed. Select the appropriate number of SonoCheck
vials and use the pictures below to choose the layout that matches the size of
the equipment. Place the SonoChecks as
indicated in an empty ultrasonic basket and place it in the ultrasonic
cleaner. Run the equipment as directed
by the manufacturer and record the test results directly on these illustrations
for troubleshooting and optimisation as needed. Refer to the equipment
manufacturers instructions to locate individual transducers if each transducer
is to be monitored

![]()
![]()




All SonoChecks should change from green to
yellow. The time needed for the color change will indicate the level of energy and
degree of cavitation provided by the ultrasonic cleaner.
- A
change slower than average will indicate a weak spot.
- A
negative result will indicate a blind spot of ultrasonic energy.
In case of an unsatisfactory result, refer to
the SonoCheck troubleshooting guide.
Optimisation necessary?
![]()
![]()
No: Yes: (repeat functional test after
optimisation)
Action
taken:
.......
.
.......
.
.... Signature: