SonoCheck trouble-shooting guide

 

 

If the SonoCheck ultrasonic cavitation monitor does not change color or if the time required to generate the color change takes longer than normal, please check the following guide:

 

Problem

 

Reason

 

Corrective action

De-gassing

 

Dissolved gasses will absorb ultrasonic energy

 

De-gas solution according to equipment manual

Water level

 

Ultrasonic energy may reflect off of the surface of the solution and change energy distribution

 

Check equipment manual for correct water level

Operating cycle time

 

Time varies with the amount of ultrasonic energy available

 

Longer operating cycles generally provide better results

Instrument load

 

Heavy instrument loading and certain materials can absorb ultrasonic energy

 

Look for weak points using the periodic functional test and check for ultrasonic absorbent material like silicone or plastics

Transducer failure

 

Transducer efficiency may decrease with age.  Individual transducers may fail while others in the equipment continue to function

 

Perform periodic functional test, placing Sonocheck monitors in each transducer location (see equipment manual)

Low energy

 

Transducer inefficiency or the ultrasonic basket may absorb too much energy

 

Check performance without basket in place.  Compare performance against another ultrasonic cleaner if available.  Call for service

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 