HemoCheck-S Instructions for use![]()
HemoCheck-S the test kit for detection of blood residue on surfaces.


1.
Open the protective pouch of the
HemoCheck-S test kit. Included are: A: Indicator-vial (transparent cap), B:
activator-vial (green cap) and C: cotton swab.
2.
Wet surfaces are swabbed with the
dry cotton swab. Dry surfaces are swabbed by moistening the swab with a drop of
clean water (Do not use chlorinated water!). Swab the sample vigorously.
3.
Open the indicator vial (A
transparent cap) and transfer the liquid into the activator-vial (B green cap).
4.
Place the sample-swab (C) into the
vial (head down into the liquid) and shake at least 5 times.
5.
Check the swab over a period of 30
seconds for a colour change to blue-green, which will indicate blood residues
on the tested surface. In the presence of large amount of blood the whole
indicator solution will change to dark blue. Record your result since small
amount of blood residue will not form a stable colour. The used test-liquid may
change to a light blue-green after several hours which does not indicate
residues.
Note: In case of jointed surgical instruments blood residues are most
common inside joints which cannot be sampled with a swab. Long swabs can be
used for checking inside cannulated instruments.

-INFORMATION-

Make it visible!
PEREG clean monitoring- the new
perspective in CSSD
Due to the high content of Peroxidases in blood an enzymatic reaction is
used for detection of blood residues.
Detection of blood residues on surfaces. Examples: Chamber walls of
WD´s, Ultrasonic cleaner, work bench, surfaces of surgical instruments. Test
for residues of blood based test soils.
The test kit can detect 0,1 µg of blood by showing a slight blue-green
spot. 1 µg of blood in the test will already give a dark blue colour.
Oxidising agent like chlorine or hypochlorite (present in some
disinfecting agents and detergents) will give a colour change too. In this case
the test cannot be used to detect blood residues.
12 single use tests for detection of blood on surfaces.
Each test kit consists of:
-
vial with indicator solution
(transparent cap)
-
activator-vial (green cap)
-
swab
STORAGE:
Store HemoCheck-S in closed pouches in a cool place 2°C- 25°C. Keep away from light and heat
SHELF LIFE: See
imprint